CARTO XP is an electroanatomical cardiac mapping system that provides 3D color-coded maps of the electrical activity of the heart; however it is expensive and it can only use a single costly magnetic catheter for each patient intervention. Our approach consists of integrating fluoroscopic and electrical data from the RF catheters into the same image so as to better guide RF ablation, shorten the duration of this procedure, increase its efficacy, and decrease hospital cost when compared to CARTO XP. We propose a method that relies on multi-view C-arm fluoroscopy image acquisition for (1) the 3D reconstruction of the anatomical structure of interest, (2) the robust temporal tracking of the tip-electrode of a mapping catheter between the diastolic and systolic phases and (3) the 2D/3D registration of color coded isochronal maps directly on the 2D fluoroscopy image that would help the clinician guide the ablation procedure much more effectively. The method has been tested on canine experimental data.
1. Introduction
The current incidence of sudden cardiac death (SCD) in the United States is between 200 000 and 250 000 cases per year, with a worldwide incidence of 4 to 5 million cases per year [1]. The most common electrophysiological mechanism leading to SCD is tachyarrhythmia such as ventricular tachycardia (VT). Severe disorders of the heart rhythm that can lead to SCD are often treated by radio-frequency (RF) catheter ablation, which consists of inserting a catheter inside the heart, near the area from which originates the abnormal cardiac electrical activity, and delivering RF currents through the catheter tip so as to ablate the arrhythmogenic area (see Figure 1(a)). The precise localization of the arrhythmogenic site and positioning of the RF catheter at that site are problematic; they can impair the efficacy of the procedure and the procedure can last several hours, especially for complex arrhythmias. To shorten the duration of RF catheter ablation and increase its efficiency during the arrhythmia, four main commercial mapping systems that provide a 3D volume and color display of the cardiac chamber and electrical activation sequence have been developed. These include: CARTO XP (Biosense-Webster, Diamond Bar, CA, USA), NavX system (St Jude Medical, St Paul, MN, USA), RPM (Boston Scientific, Natick, MA, USA), and LocaLisa (Medtronic, Minneapolis, MN, USA). All these systems including purchase of system-specific catheters are costly for the modern day hospital.
Figure 1. (a) Ablation catheter and RF currents during energy delivery. (b) Monoplane C-arm fluoroscopic X-ray system to guide RF catheter ablation.
Other approaches proposed by researchers to guide RF ablation therapy employ multimodal image fusion. These include the visualization of an optically tracked catheter by making use of magnetic resonance imaging (MRI) [2, 3], the combination of MRI and fluoroscopy [4], ultrasound imaging of the ablation catheter [5], the combination of ultrasound and preoperative computer tomography (CT) [6], or preoperative imaging (CT/MRI) for ablation planning [7, 8]. However, all these approaches omit incorporating electrophysiological data that is crucial for the clinician when locating the arrhythmogenic site.
Recently, efforts have been made in attempting to develop a more affordable fluoroscopic navigation system by obtaining local electrical activation times from a roving ablation catheter whose positions are computed from a single image using a monoplane C-arm fluoroscope. However, this single image method yielded inaccurate 3D reconstructions as depth estimations were about 10 mm, thus motivating the need of using multiview geometry for accurate 3D reconstruction. Research objectives were aimed at emulating the format of the CARTO XP system and focusing on treating ventricular tachycardias [9]. With the CARTO XP nonfluoroscopic mapping approach, the patient is positioned over a tripod emitting three electromagnetic waves at unique frequencies. Each beam is registered by one of three specifically tuned coils embedded in the mapping catheter tip to specify location in 3D space, when the catheter tip is considered against a reference catheter. The catheter location and electrograms are recorded and reconstructed in realtime and presented as a 3D geometrical map color coded with the electrophysiological information. The arrhythmia must remain the same during the long point-by-point mapping procedure.
The emphasis of this paper is a continuation of the work in [9], but with a focus on multiview vision techniques. The long-term objective continues
to develop a system similar to the functionality of the CARTO XP technology; albeit with the following advantages: (a) being more affordable by making
use of cost-effective catheters which increase the number of available types and shapes
of catheters used during the cardiac ablation procedure, (b) use of common monoplane
C-arm X-ray fluoroscopes (see Figure 1(b)) compared to expensive new mapping equipment (
300,000$/system,
5,000$/catheter) that need to be purchased by hospitals, and (c) that 2D/3D registration
using only X-ray fluoroscopy is possible in order to superimpose a translucent image
of the cardiac activation map directly over 2D C-arm images.
The contributions of this paper are as follows: (i) we provide a complete analysis on two-view 3D reconstruction of the tip-electrode of a mapping catheter inserted in the left ventricle of the heart, (ii) we propose an automatic algorithm that extracts the fluoroscopy image depicting the diastolic image phase, (iii) we automatically filter and track the tip-electrode in a sequence of images using a 2D/2D registration algorithm initialized by clicking once on the tip electrode seen in the diastolic image, and (iv) we provide a detailed report on the feasibility of using our multiview 3D methodology to guide VT catheter ablation by presenting an experimental procedure and results on canine data.
2. Methodology
The overall workflow is summarized as follows. Once C-arm data acquisition is achieved, extracting relevant images is necessary in order to minimize motion artifacts before reconstruction in three dimensions. Thus, an image extraction algorithm is first presented allowing us to select images depicting important phases of the cardiac cycle. Second, we describe a single-click algorithm coupled with a new image processing scheme allowing for automatic temporal tracking of the tip electrode. We conclude the section by recalling the mathematics of multiview geometry and propose a fitting technique for 3D reconstruction purposes.
2.1. Fluoroscopic Image Analysis
As an alternative to electrocardiogram gating and to minimize motion blur in the C-arm images acquired, we developed the following algorithm to select the frame for each mapping site. The number of frames making up each sequence varied depending on how long the technician stepped on the C-arm foot pedal. In general, each sequence contained up to 200 frames. By calculating the standard deviation of the differences between all the pixels of two successive frames, we were able to visually determine that the frame showing the smallest standard deviation also had the least motion blur (see Figure 2). This frame represented the diastolic phase of the cardiac cycle. The maximum standard deviation occurred during the systolic phase.
Figure 2. Standard deviation of the differences between successive frames during a one-second
sequence (60 frames). Shown here is a sample curve depicting the image extraction algorithm and the diastole
image frame number.
2.2. Automatic Tracking of Tip-Electrode Using Single-Click Initialization
Locating the tip-electrode of the mapping catheter is crucial since it will come into contact with the arrhythmogenic site for ablation purposes. Therefore, extracting 2D coordinates of the tip electrode is essential. However, this is not an easy task when attempting to develop an automatic algorithm. To our knowledge, the work in [10] presents the most recent work focusing on this task. The authors present a steerable tensor voting filter to extract the tip-electrode of ablation catheters and report results of approximately 70% accuracy for electrode detection in extremely noisy images. Nevertheless, the algorithm requires defining 3 tensor voting field parameters that require adjustment. The results are not clinically viable and manual selection is still necessary to collect all the tip-electrode coordinates in the C-arm images. At this stage, we thought it best to preserve a smooth workflow by manually selecting only the tip electrode visible in each of the diastole images.
2.2.1. Three-Step Filter to Catalyze Temporal Tracking
At each user-click of the tip-electrode, automatic tracking commences between the
diastolic and systolic image frames determined in Section 2.1. Two cropped window
sizes are first defined: (i) a
template window surrounding the tip electrode center, and a
search window around the template. For both cropped windows, we apply the following 3-step preprocessing filter.
A homomorphic filter is used first to denoise the fluoroscopic image [11]. The homomorphic filter decreases the contribution made by the low frequencies and amplifies the contribution of high frequencies. The result is simultaneous dynamic range compression and contrast enhancement. The homomorphic filter is given by
(1)with
and
. The coefficient c controls the sharpness of the slope at the transition between high and low frequencies;
whereas
is a constant that controls the shape of the filter and
is the distance in pixels from the origin of the filter.
Anisotropic systems are those that exhibit a preferential spreading direction while isotropic systems are those that have no preferences. The Perona-Malik anisotropic diffusion [12] method was implemented here in order to reduce noise and texture from the image, as well as to preserve and enhance structures. The diffusion equation is given by
(2)where
is the input image and
is the diffusion coefficient that controls the degree of smoothing at each image
pixel. The diffusion coefficient is a monotonically decreasing function of the image
gradient magnitude. It allows for locally adaptive diffusion strengths; edges are
selectively smoothed or enhanced based on the evaluation of the diffusion function.
Although any monotonically decreasing continuous function of the gradient would suffice
as a diffusion function, we use the following diffusion coefficient:
(3)
is referred to as the diffusion constant or the flow constant. The greatest flow is produced when the image gradient magnitude is close to the
value of
. Therefore, by choosing
to correspond to gradient magnitudes produced by noise, the diffusion process can
thus be used to reduce noise in images.
Morphological filtering was applied as a final image processing step in order to eliminate background elements around the primary coronary arteries. The structuring element consists of a pattern specified as the coordinates of a number of discrete points relative to a defined origin. Normally, Cartesian coordinates are a convenient way of representing an element as a small image on a rectangular grid. We chose a disk structuring element having a radius of a few pixels, since the contours of the tip-electrode of the mapping catheter can be modeled as a disk (see Figure 3). The structuring element will suppress the background (black) and enhance the arteries (grayscale). When a morphological operation is carried out, the origin of the structuring element is typically translated to each pixel position in the image in turn, and then the points within the translated structuring element are compared with the underlying image pixel values.
Figure 3. The tip-electrode of an ablation catheter approximated as a disk (circle).
2.2.2. Temporal Tracking via 2D/2D Registration
Our method builds on the following intuition. As we are dealing with the same imaging modality during registration, then the electrode tip carries enough distinctive information for intensity-based registration to hone in the correct transformation (i.e., displacement) between image frames. Also, since the electrode structure is solid then a rigid-based 2D/2D registration scheme should suffice.
We implemented the Normalized Correlation (NC) metric. Pixel values are taken from the search window; their positions are mapped to the template image in order to find a match, or superposition of the two. The correlation is normalized by the autocorrelations of both the images [13]. Let Img1 and Img2 be the search window and template image, respectively. NC computes pixel-wise cross-correlation and normalizes it by the square root of the autocorrelation of the images:
(4)where
and
are the ith pixels in two images, respectively, and N is the number of pixels considered. The
factor is used to optimize the metric when its minimum is reached, say at minus one.
Misalignment between the images results in small measurement values. This metric produces
a cost function with sharp peaks and well-defined minima. The number of spatial samples
used here is empirically set at 50.
The transform being updated is a simple Euler 2D transform which is composed of a plane rotation and a two-dimensional translation. The rotation is applied first, followed by the translation. Optimization was performed using the Levenberg-Marquardt algorithm and maximum iterations were set to a maximum of 100. Seeing how both images are relatively small to begin with, convergence should be relatively instant.
Figure 4. Workflow to acquire tip electrode coordinates in successive 2D images between the
diastolic and systolic phase. A single click initialization launches a robust tracking routine.
The next position of the tip electrode is obtained by calculating optimal transformation that locks both images over each other. The centroid is extracted and serves as 2D coordinates for the tip electrode in the image following the diastolic frame. Here, template and search window images are extracted again and registration is executed. This routine is performed until the systolic image (see Figure 4). At this stage we should have the estimated 2D coordinates of the tip electrodes in all images between diastole and systole.
2.3. C-Arm Fluoroscopy Geometry
If we define a three-dimensional point
in the world coordinate system (see Figure 5), then its 2D projection in an image,
, is achieved by constructing a projection matrix:
Figure 5. The perspective camera model. Any 3D world point can be projected onto a 2D plane and its coordinates would be
(
) pixels.
(5)The intrinsic matrix of size [
] contains the pixel coordinates of the image center, also known as the principal
point
, the scaling factor
, which defines the number of pixels per unit distance in image coordinates, and the
source-to-image distance SID, also known as the focal length, f of the C-arm (in meters). The extrinsic matrix of size
is identified by the transformation needed to align the world coordinate system to
the camera coordinate system. This means that a translation vector,
, and a rotation matrix,
, need to be found in order to align the corresponding axis of the two reference frames.
Imaging parameters such as focal length, primary and secondary rotation angles image
pixel spacing and size are obtained from the accompanied DICOM image files. This allows
us to have a close enough estimate of the C-arm pose for a particular viewing angle.
Spatial reconstructions from biplane geometry evolved as important tools for morphological analyses of vessels or catheters in both cardiology and neurology domains [14–19]. From the known imaging geometry and using the point correspondences established by the epipolar constraint, any point visible in both projections can be spatially reconstructed by retracing the projection rays back to the point of their intersection. This can be achieved after successful calibration of the C-arm geometry. Since the reconstructed rays often miss this point due to slight reconstruction or calibration errors, usually their closest location is approximated. This is known as a stereo triangulation procedure (see Figure 6). Using the manual clicked tip electrode center coordinates, as well as the initial parameters of the C-arm gantry, we optimized the final 3D coordinates using the triangulation method. Details are omitted here for the sake of brevity.
Figure 6. Stereo triangulation setup for the 3D reconstruction of corresponding points located
in biplane images. Since both backprojected rays from camera
and camera
do not intersect, a solution is found by estimating a 3D point, P, which is half
way between the two rays.
2.4. Convex (Quick) Hull Algorithm
The tip-electrode coordinates in the monoplane images are collected and used to form
a 3D volume using a convex hull algorithm. The problem involves finding the smallest
convex polygon containing all the points of
, given a set
of
points in multidimensional space [20]. The algorithm can be described briefly for 2D points as seen in Figure 7: Select the farthest points (left and right) from the data set (2D) and draw a line
between them. We examine a set of points lying on the same side of the line (CD).
In this set, select the point E that is located furthest away from the line CD. This
point will also belong to the convex hull, since it cannot be included in a triangle.
Moreover, we can remove all the points inside the triangle (CDE) and split the remaining
points into two subsets: one with the points on the left of line (CE), and the other
set containing points on the right of line (CE). The iterative process is repeated
in these two subsets. Once the outer shell of the hull is established, we can analyze
the set of data points located under the line segment (CD) in the same manner. Similar
procedure can be expanded to the 3D scenario [14]. The convex hull will have an empty interior and is the method of choice in this
paper. It is quick to implement and represents well the endocardial surface of cardiac
chambers. Further, it was thought to be a good initial step for verifying the feasibility
of our approach.
Figure 7. Convex Hull from a set of unorganized points.
3. Experiments and Results
3.1. Mongrel Dog Experiment
We begin this section by recalling that our primary objective was to develop and test a mapping technique based on multiview measurements of electrode sites and on the reconstruction of the cardiac chamber of interest. Here, the reconstruction uses a full perspective projection model coupled with surface fitting Convex Hull algorithm. Our overall methodology mimics the CARTO XP technology but only requires cost-effective standard catheters and a monoplane C-arm fluoroscope.
An animal experiment was carried out at Sacré Coeur Hospital, Montreal, Canada. Our
experimental protocol met the institutional requirements of animal experimentation.
A mongrel dog was anesthetized and laid on its right side on a fluoroscopy table (Integris
Allura, Philips Inc.). A reference catheter and a pacing catheter were inserted into
the right ventricle, close to the septal wall. The role of the reference catheter
was to define an origin for our 3D coordinate system that is not affected by the displacement
of the heart due to respiration or by the displacement of the animal or the X-ray
system. The role of the pacing catheter was to produce a simple electrical activation
sequence so as to validate the isochronal maps. Finally, a standard 7-French RF ablation
catheter was inserted from the femoral vein into the left ventricle of the dog. During
the course of the experiment, this mapping catheter was moved to 20 different sites
(point-by-point) within the ventricle in order to obtain electrical and geometrical
data from sufficient sites to map the activation sequence. Electrograms were recorded
with the mapping catheter and the reference catheter using a multichannel analog-to-digital
(A/D) converter card (PCI DAS-1001, Omega Engineering Inc.). The electrograms were
saved in text file format. Specifications for the acquisition software were: 1000
samples/second, 1000 Hz bandpass, and 0.05 Hz high-pass and 450 Hz low-pass frequencies.
Local activation time was measured as the difference in the times of the fastest negative
deflections (dV/dt) seen in the two electrograms recorded with the reference catheter
and the mapping catheter. The fluoroscopic image acquisition rate was set at 60 frames/second
and images were recorded for about 2 seconds at the end of the expiration. The total
fluoroscopy exposure time was thus 80 seconds which yielded a radiation dose of about
25 mGy to the mongrel dog. As illustrated in Figure 8, the monoplane fluoroscopic C-arm was rotated by
to acquire two images for each mapping site: a left lateral view (the C-arm in a
vertical position) and a posterior view (the C-arm in a horizontal position). Images
were recorded, in DICOM format, with a
pixel resolution. In the end, the data collected totaled twenty posterior view images,
twenty left lateral view images, and the twenty electrograms. The MatLab (R2008a)
software was used for all methods developed in this paper for prototyping purposes.
Figure 8. (a) Typical monoplane C-arm fluoroscope and coordinate system centered at the tip
of the reference catheter electrode. Biplane images are obtained by rotating the C-arm at two perpendicular positions.
(b) Sample diastolic images for posterior and left lateral, respectively.
3.2. Temporal Tracking Results
The forty datasets were analyzed using our image extraction algorithm as outlined in Section 2.1. Thus, a total of forty diastole and systolic images were extracted from the DICOM datasets, plus additional images between the two cardiac phases. At this point, we manually clicked the tip electrodes of both the reference and mapping catheters in all diastole images and recorded them for the multiview 3D reconstruction method.
We simultaneously launched the automatic tracking algorithm once the mapping electrode
was manually selected. Thetemplate and search box were instantiated followed by the application of the 3-step filter as seen in Figure
9. Registration was launched until the systolic image was reached. Estimated center
results of the tip electrode were compared to an expert user click on the tip electrode
for all images between diastole and systole. From Table 1, the total number of images analyzed for the posterior and left lateral datasets
was 181 and 189, respectively. The average tip-electrode location error was
pixels and
pixels in the
and
directions for the posterior images. On the other hand, results improved for the
left lateral view, as the average tip-electrode location error was
pixels and
pixels in the
and
directions, respectively. The excellent image quality for the left lateral images
reflects the more precise tracking results.
Table 1. Temporal tracking results for both posterior and left lateral datasets.
Figure 9. (a) Template and search boxes defined around the centroid of the tip-electrode in
the posterior and left lateral views respectively. (b) The final images from the 3-step filter applied to the template and search boxes
respectively. Note the suppression of background in the images.
In fact, three of the twenty posterior datasets contributed to the large maximum registration errors as seen in Table 1 as image quality and motion blur were ever present. A sample view of the tracking results is shown in Figure 10. The white dotted box shows the recovered positions after registration.
Figure 10. (Top two rows) Sample tracking results for the posterior view between diastole and systole. Random middle images are shown between the two cardiac phases. (Bottom two rows) Example images for the left lateral view.
3.3. Multiview (Biplane) 3D Reconstruction Results
We subtracted the coordinates of the reference catheter for each mapping site so as to compensate for respiration and motion artifacts. The 3D coordinates of the tip-electrodes were optimized and recovered by the triangulation algorithm. We fitted the cloud of points using the 3D Convex Hull algorithm volume shown in Figure 11. After creating the 3D convex hull, the electrical activation times measured at each node were linearly interpolated over the surface of this volume and displayed as a color shift, using red for the earliest activation times and blue for the latest. The volume reflects the left ventricle of the mongrel dog.
Figure 11. Three-dimensional convex hull of left ventricle obtained using the triangulation and
perspective camera model. The (
) symbol represents the tip electrode coordinates of the mapping catheter which was
moved to 20 different locations. Activation times were linearly interpolated and color
coded with a red color signifying earliest activation time.
So as to plot only the triangles that can be seen from a certain point of view, the
normal to the surface of each triangle was calculated (the normal points outward),
and if its dot product with the observation point vector was positive, the triangle
was retained. For example, for the posterior view (
plane), we retained all the nodes associated with a triangle whose normal component
in the
direction is positive. Similarly, for the left lateral (
) plane, we retained the nodes associated with a triangle whose normal component in
the
direction is negative. The fusion between the 2D isochronal activation map and the
fluoroscopic image was performed in a 6-step process (see Figure 12), as outlined in the following.
Figure 12. (a) Isochronal map superimposed on the diastolic images of the first four datasets
for the left view and (b) posterior view.
(1)The nodes visible from the given point of view are identified using the Convex
Hull data, and their 2D coordinates are retained (e.g., for the posterior view, the
nodes are in the
plane).
(2)The activation times of the visible nodes are positioned in a matrix, at the 2D coordinate locations of the nodes.
(3)The Delaunay method is used to organize the set of visible points as a set of triangles and cubic interpolation is applied to interpolate the activation times inside each triangle.
(4)In order to fuse the isochronal map to the [
] size fluoroscopic image, the borders of the above matrix are completed with zeros
to achieve the same size as the image.
(5)The values of the [
] activation time matrix are scaled so that the maximum value is equal to 64 (which
correspond to blue) and the minimum value is equal to 0 (which correspond to red).
(6)The scaled activation time matrix is used to create an RGB color matrix (using a red-to-blue color map), which is finally plotted over the fluoroscopic image with a 30% transparency ratio (30% for the color isochronal map and 70% for the grayscale fluoroscopic image).
From Figure 12, earliest activation times are clearly closest to the stimulation catheter (near the reference catheter). The posterior view also shows that the isochronal map followed the contour of the mapping catheter, which corresponds to the contour of the left ventricular cavity.
3.3.1. Present Method Advantages/Limitations versus the CARTO XP Technology
Since the CARTO XP nonfluoroscopic technology provides contact-based sequential acquisition of endocardial
signals and reconstruction of 3D electroanatomical maps, this can be time consuming
in terms of acquisition. However, this factor is operator dependent, and in experienced
hands final geometry of a cardiac chamber can be obtained within 30 minutes [21]. More important, the instability of the catheter used for timing intracardiac activation
and major patient movements relative to the location pad may render the entire map
inaccurate for subsequent use, requiring the construction of a whole new map. For
established indications, the major limitation for broader use of electroanatomical
mapping in catheter ablation procedures at the present time is the associated cost.
This is further confounded by the high cost of the mapping catheter. In terms of typical
ventricular tachycardia ablation procedures, authors in [22] indicate total average procedure time of
minutes with fluoroscopy time of
minutes.
The major drawback of our methodology is the use of fluoroscopy in order to reconstruct
an anatomical 3D structure when compared to CARTO XP. Data acquisition took approximately 30–40 minutes, with about an additional 10–15
minutes of manual selection of the tip-electrodes, filtering, and temporal tracking
for all image frames between diastole and systole. Also, the sequential measurements
of the horizontal and vertical fluoroscopic images were not as accurate as simultaneous
measurements with a true biplane fluoroscope as the vertical position of the reference
catheter showed a difference of
pixels in the twenty posterior and left lateral images. This relatively small difference
can be explained by motion artifacts (respiration, heartbeat, subject displacement)
and fluoroscope misalignment. However, we provide an added advantage by introducing
2D isochronal activation maps superimposed over the corresponding fluoroscopic images
in the diastolic phase. These isochronal maps accurately depicted the progression
of the electrical activation away from the pacing catheter. Furthermore, our entire
methodology is cost efficient and this is important as the availability of expensive
systems such as CARTO XP is still limited in developing countries [21].
4. Conclusion
We introduced a multiview method of a fluoroscopic navigation system to guide RF catheter
ablation of cardiac arrhythmias. This initial prototype allowed us to: (
) localize in 3D the mapping catheter tip-electrode from two-view fluoroscopic images;
(
) automatically extract important images of the cardiac phase, (
) measure the local activation times on the electrograms recorded with the catheter
at multiple sites, and (
) superimpose over the fluoroscopic images, isochronal maps depicting the electrical
activation sequence from which the cardiologist could precisely localize the arrhythmogenic
site with respect to the RF catheter. As a plus, we present a single-click tracking
algorithm that could estimate the locations of the tip-electrode in images following
diastole. Our overall methodology strikes a balance between efficiency and robustness
of a common multiview vision problem applied to medicine. Nevertheless, the true effectiveness
of our method will be determined on patient data validation and experimentation.
Acknowledgment
The author would like to thank Dr. Pierre Savard, from École Polytechnique, Canada, for providing the image datasets and valuable suggestions.
References
-
EC Stecker, C Vickers, J Waltz, et al. Population-based analysis of sudden cardiac death with and without left ventricular systolic dysfunction: two-year findings from the Oregon sudden unexpected death study. Journal of the American College of Cardiology 47(6), 1161–1166 (2006). PubMed Abstract | Publisher Full Text
-
AC Lardo, H Halperin, C Yeung, P Jumrussirikul, E Atalar, E McVeigh, Magnetic resonance guided radiofrequency ablation: creation and visualization of cardiac lesions. Proceedings 1st International Conference on Medical Image Computing and Computer-Assisted Interventation (MICCAI '98), October 1998, Cambridge, Mass, USA, 189–196
-
AC Lardo, ER McVeigh, P Jumrussirikul, et al. Visualization and temporal/spatial characterization of cardiac radiofrequency ablation lesions using magnetic resonance imaging. Circulation 102(6), 698–705 (2000). PubMed Abstract | Publisher Full Text
-
E Chu, AP Fitzpatrick, MC Chin, K Sudhir, PG Yock, MD Lesh, Radiofrequency catheter ablation guided by intracardiac echocardiography. Circulation 89(3), 1301–1305 (1994). PubMed Abstract | Publisher Full Text
-
Y Sun, S Kadoury, Y Li, et al. Image guidance of intracardiac ultrasound with fusion of pre-operative images. Proceedings of the 10th International Conference on Medical Imaging and Computer-Assisted Intervention (MICCAI '07), October-November 2007, Brisbane, Australia, Lecture Notes in Computer Science 4791, 60–67
-
R Razavi, DLG Hill, SF Keevil, et al. Cardiac catheterisation guided by MRI in children and adults with congenital heart disease. The Lancet 362(9399), 1877–1882 (2003). Publisher Full Text
-
R Kato, L Lickfett, G Meininger, et al. Pulmonary vein anatomy in patients undergoing catheter ablation of atrial fibrillation: lessons learned by use of magnetic resonance imaging. Circulation 107(15), 2004–2010 (2003). PubMed Abstract | Publisher Full Text
-
JM Lacomis, W Wigginton, C Fuhrman, D Schwartzman, DR Armfield, KM Pealer, Multi-detector row CT of the left atrium and pulmonary veins before radio-frequency catheter ablation for atrial fibrillation. Radiographics 23, S35–S48 (2003). PubMed Abstract | Publisher Full Text
-
P Fallavollita, P Savard, G Sierra, Fluoroscopic navigation to guide RF catheter ablation of cardiac arrhythmias. Proceedings of the 26th Annual International Conference of the IEEE Engineering in Medicine and Biology (EMBC '04), September 2004, San Francisco, Calif, USA 3, 1929–1932
-
E Franken, P Rongen, M Van Almsick, B Ter Haar Romeny, Detection of electrophysiology catheters in noisy fluoroscopy images. Proceedings of the 9th International Conference on Medical Image Computing and Computer-Assisted Intervention (MICCAI '06), October 2006, Copenhagen, Denmark, Lecture Notes in Computer Science 4191, 25–32 PubMed Abstract
-
VI Ponomarev, AB Pogrebniak, Image enhancement by homomorphic filters. Applications of Digital Image Processing XVIII, July 1995, San Diego, Calif, USA, Proceedings of SPIE 2564, 153–159
-
P Perona, J Malik, Scale-space and edge detection using anisotropic diffusion. IEEE Transactions on Pattern Analysis and Machine Intelligence 12(7), 629–639 (1990). Publisher Full Text
-
TS Yoo, MJ Ackerman, WE Lorensen, et al. Engineering and algorithm design for an image processing Api: a technical report on ITK—the Insight Toolkit. Studies in Health Technology and Informatics 85, 586–592 (2002). PubMed Abstract | Publisher Full Text
-
Z Zhang, Determining the epipolar geometry and its uncertainty: a review. International Journal of Computer Vision 27(2), 161–195 (1998). Publisher Full Text
-
PHS Torr, DW Murray, The development and comparison of robust methods for estimating the fundamental matrix. International Journal of Computer Vision 24(3), 271–300 (1997). Publisher Full Text
-
G Shechter, Respiratory motion of the heart: implications for magnetic resonance coronary angiography, Ph.D. dissertation
-
K Sprague, M Drangova, G Lehmann, et al. Coronary x-ray angiographic reconstruction and image orientation. Medical Physics 33(3), 707–718 (2006). PubMed Abstract | Publisher Full Text
-
N Sang, W Peng, H Li, Z Zhang, T Zhang, 3D reconstruction of the coronary tree from two X-ray angiographic views. Medical Imaging 2006: Image Processing, February 2006, San Diego, Calif, USA, Proceedings of SPIE 6144
-
C Blondel, G Malandain, R Vaillant, N Ayache, Reconstruction of coronary arteries from a single rotational X-ray projection sequence. IEEE Transactions on Medical Imaging 25(5), 653–663 (2006). PubMed Abstract | Publisher Full Text
-
CB Barber, DP Dobkin, H Huhdanpaa, The quickhull algorithm for convex hulls. ACM Transactions on Mathematical Software 22(4), 469–483 (1996). Publisher Full Text
-
F Duru, CARTO three-dimensional non-fluoroscopic electroanatomic mapping for catheter ablation of arrhythmias: a useful tool or an expensive toy for the electrophysiologist? Anadolu Kardiyoloji Dergisi 2(4), 330–337 (2002). PubMed Abstract | Publisher Full Text
-
Y-G Li, Q-S Wang, G Grönefeld, et al. Refinement of CARTO-guided substrate modification in patients with ventricular tachycardia after myocardial infarction. Chinese Medical Journal 121(2), 122–127 (2008). PubMed Abstract | Publisher Full Text